Non-Invasive Blood Pressure
Technique: Oscillometric measure upon inflation
Measurement Time: <40 sec typical; 0 to 300mmHg; standard adult cuff
Automatic Measurement Cycles: 1 - 60 min; 2 - 4
hr; STAT
Inflation Pressure Range: 30 to 300mmHg (Adult); 30 to 150mmHg (Pedicatric);
20 to 150mmHg (Neonate)
Resolution: 1mmHg
Clinical Accuracy: SP-10:2002
Transducer
Accuracy: ±2mmHg or ±2% of reading, whichever is greater, over full
range
STAT mode: 5 min of consecutive readings
Pulse Oximetry
Range: 1% to 99%
Resolution: 1%
Accuracy (DOX only): 70% to 99% range: ±2%; 50% to 69% range: ±3%; <50%:
unspecified Accuracy (OxiMax only): 70% to 99% range: ±2%; <70%:
unspecified
Method: Dual wavelength LED
Audio: Tone with each detected pulse; pitch varies with saturation
Heart (Pulse) Rate
Range: 20 to 300bpm
Accuracy (DOX only): Greater of ±1bpm or ±1%, of reading unspecified for <30
or > 240 bpm Accuracy (OxiMax only): Greater of ±1bpm or ±1%, of reading unspecified for <30
or > 240 bpm
Update Time: Every sec.
Resolution: 1bpm
Pulse Volume: 10 levels or Off
Temperature
Type: FILAC FasTemp
Method: Predictive or Continuous
Channels: 1
Range: 30.0º to 43.0º C
Accuracy: ±0.2º C for 30.0º to 36.9º C; ±0.1º C for 37.0º to 39.0º C;
±0.2º C for 39.1º to 43.0º C
Resolution: 0.1º F or 0.1º C
Predictive Response Time: Oral: less than 10 seconds;
Axial: less than 10 seconds;
Rectal: less than 10 seconds
Probes/Wells: Oral/Axial FILAC FasTemp
Rectal FILAC FasTemp (optional)
Isolation Well, Oral/Axial
Isolation Well, Rectal (optional)
Alarms
Indication: Audible; visual
Volume: 10 levels or Off
Silence: Yes; 2 min. or permanent
Trend Memory
Memory: Last 24 hrs; SpO2, pulse rate, NIBP,
temperature in Single Patient or Multi-Patient Report Types with
selectable formats.
Displays
Type: LED
Parameters: NIBP, SpO2; temperature, pulse rate; pulse bar
Status: Alarm Silence; Sensor; Low Battery; External Power; Patient Size
Text: Alphanumeric LCD
Languages: Eng., Fr., Ger., It., Port., Span.
System Output
Com Port: Serial, RS 232-compatible
Printer
Type: Thermal array
Tabular Mode: 1 - 60 min intervals; 2 - 24 hr
intervals; Auto on BP; Auto on Alarm, Off
Mechanical
Weight: Base unit with battery: 6.95 lb; 3.15kg
Size: Base unit: 8.7" (H) x 7.1" (W) x 6.5" (D); 22cm
(H) x 18cm (W) x 16.5cm (D)
Power Requirements
Voltage: 50 to 60 Hz: 100 to 120 VAC
Battery: Sealed lead acid; life: 8 hr; recharge
time: 6 hr
Environmental
Operating Temperature: 32º to 113º F; 0º to 45º C
Storage Temperature: 23º to 122º F; -5º to 50º C
Operating Humidity: 15% to 90%, non-condensing
Classification
Type of Protection: Class I equipment
Degree of Protection: Type CF defibrillator-proof equipment
Protection against Ingress: IPX1 rating, Drip-Proof Equipment
CE compliant. ETL Certified.
Specifications subject to change without notice.
|