Dimensions
9.5”H x 13”W x 8”L maximum
Weight
13 lbs. Maximum without batteries
Power Requirements
Power consumption: 15 watts typical. Voltage input: 100, 120, 220, 240 VAC ± 10%. Mains frequency: 47 to 65 Hz.
ECG
Leads: I, II, III, aVR, aVL, aVF, V. Display sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ±10%. Frequency response to screen: ESU filter out: 0.5 to 40 Hz, -3db max @ 1 mVpp input. ESU filter in: 0.5 to 20 Hz –3db max @ 1 mVpp input
Heart Rate Meter
Range: 30-300 bpm adult/child. 30-350 bpm neonatal. Accuracy: 30-250 bpm ± 3bpm or 3%, whichever is greater. Pacer rejection: rejects all pulses of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with no tail. Rejects all pulses ±2.0 mV to ±700 mV and duration 0.1 ms to 2 ms with 100ms T.C. tail <0.8 mV or 25 ms T.C. tail < 0.4 mV. Display update: 2 seconds ± 0.1 sec
NIBP
Technique: oscillometric. Systolic range: adult/ped mode: 50 –250 mmHg, neonate mode: 15-150 mmHg. Diastolic range: adult/ped mode: 30-200 mmHg, neonate mode: 10-150 mmHg. Systolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 3 mmHg, standard deviation less than 13 mmHg (neonatal). Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal). Heart rate meter range: 30-250 bpm. Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater.
Pulse Oximetry
Accuracy (1standard deviation): 70-100%, ±2%; 60-69%, ±4%; 0-59%, unspecified. Response time: 13 seconds to 95% of final step change % SpO2 value from 60-95% at 72 bpm
Recorder
Speed: 50, 25, 12.5, 6.25, 3.125 mm/sec ±5%. Note: 6.25 and 3.125 are for respiration only. Chart size: 50 mm ±2mm
Battery Requirements
2 sealed lead batteries. Voltage: 12 VDC. Capacity: 2.3 AH each. Run time: 2 hours fully charged battery. Recharge time: 16 hours maximum. Low battery warning: 11.6 ±3.5% . Low battery cut off: 11.0 ±2.8%.