Dimensions
Height: 27.4 cm. Depth: 24.3 cm. Width: 29.3 cm. Weight: 5.7 kg.
Display
Size: 26 cm (diagonal). Type: active-matrix color TFT. Resolution: 640 by 480 pixels. Number of traces: 6 maximum. Number of seconds/trace: 4.9 at 25 mm/sec. Sweep speed: 6.25, 12.5, 25 mm/sec (with erase bar). Information window: displays non-real-time information without obstructing the display of real-time information. Display organization: prioritized by parameter.
Controls
Trim Knob® control. Five hard keys: silence alarm, graph go/stop, function (zero all) and power on/off.
Alarms
Categories: patient status and system status. Priority: 4 levels – crisis, warning, advisory, and message. Notification: audible and visual. Setting: default and individual. Silencing: 1 minute, current alarm only. Pause: 5 minutes in adult ICU mode, 3 minutes in neonatal ICU mode, and 5, 15, or permanent pause in OR mode. Volume: default 70%, 70 dB measured at 1 meter.
ECG
Standard leads available: I, II, III, V, aVR, aVL, and aVF. 10 lead wire cable: I, II, III, V, aVR, aVF, V2, V3, V4, V5, V6. Leads analyzed simultaneously: I, II, III, and V (multi-lead mode). Lead fail: identifies failed lead. Alarms: user-selectable upper and lower heart rate limits.
Input Specifications
Voltage range: ±0.5 mV to ±5 mV. Signal width: 40 ms to 120 ms (Q to S). Heart rate range: 30 to 300 bpm. Input impedance: common mode > 10 Mat 50/60 Hz, differential > 2.5 Mfrom dc to 60 Hz. Common mode rejection: 90 dB minimum at 50 or 60 Hz
Respiration
Measurement technique: impedance variation detection. Range respiration rate: 0 – 200 breaths per minute. Base impedance: 100 – 1000 Wat 52.6 kHz. Detection sensitivity: 0.4 to 10 Wvariation. Waveform display bandwidth: 0.1 to 1.8 Hz (-3 dB). Alarms: user-selectable upper and lower respiration limits, and user-selectable apnea limit.
Temperature
Number of channels: 2. Input specifications – probe type: YSI series 400 or 700, temperature range: 0°C to 45°C (32°F to 113°F), resolution: ±0.1°C. Output specifications - parameters displayed: T1, T2, accuracy: ±0.1°C for YSI series 4000 probes, ±0.3°C for YSI series 7000 probes, independent of source. Alarms: user-selectable upper and lower limits for T1 and T2.
Invasive Blood Pressure
Number of channels: 2 (optional). Transducer sites: arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial, intracranial, and special. Transducer requirements: 5 Vdc ±0.1% excitation voltage. Transducer output: 5 µV/V/mmHg. Input specifications – range: -25 mmHg to 300 mmHg, offset: ±150 mmHg. Output specifications – frequency response: dc to 50 Hz (-0/+2 Hz). Zero balance range: ±150mmHg. Zero balance accuracy: ±1mmHg. Zero balance drift: ±1mmHg over 24 hours. Accuracy: ±2% or ±1mmHg, whichever is greater. Alarms: user-selectable upper and lower limits for systolic, diastolic, and mean pressures.
NIBP
Measurement technique: oscillometric. Displayed parameters; systolic, diastolic, and mean pressures. Systolic: 30-275 mmHg adult, 30-235 mmHg pediatric, and 30-135 mmHg neonate. Diastolic: 10-220 mmHg adult, 10-220 mmHg pediatric, and 10-110 mmHg neonate. MAP: 20-260 mmHg adult, 20-260 mmHg pediatric, and 20-125 mmHg neonate. Pulse rate, as displayed in tabular trends: 30-200 bpm adult, 30-200 bpm pediatric, and 30-200 bpm neonate
Pulse Oximetry
Parameters monitored: arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR). SpO2 range: 50 –100%. PPR range: 30-300 beats per minute. SpO2 accuracy: ±2% (70-100%) and ±3% (50-69%). PPR accuracy: ±3 beats per minute. Alarms: user-selectable upper and lower limits for SpO2 and PPR.
CO2Information displayed: inspired and expired CO2 concentrations in %, mmHg, or kPa, respiratory rate, and continuous CO2 waveform. Measurement ranges: PiCO2/FiCO2: 0 to 100 mmHg/0to13%. Pe CO2/ Fe CO2: 0 to 100 mmHg/0 to 13%; RR: 0 to 120 breaths/min. Accuracy: CO2: 5% of reading or ±2 mmHg, whichever is greater. Display resolution: 1 mmHg. Rise time: less than 60 msec. Respiration rate accuracy: ± breath/min