Test Types
Forced Vital Capacity (FVC), Flow Volume Loop (FLV), SVC, Maximum Voluntary Ventilation (MVV), OSHA Cotton Dust, Disability
Parameters
FEV1, FVC, (or FEV6), FEV1/FVC, (or FEV1/FEV6), MVV, FEF 25%-75%, PEF, FET, FIVC, PIF, variability, pre-post % change, VC, IRV, ERV, VT, IC, FEV1/VC
Predicted Normals
NHANES III, Knudson 76, Knudson 83, Morris, Crapo, Dockery, Hsu
Languages
English, Spanish and French
Accuracy
Volume: +- 3% or 0.050 L, whichever is greater
Flow: +- 5% or 0.200 L/s, whichever is greater (except PEF)
PEF: +- 10% or 0.400 L/s, whichever is greater
Range
Volume: 0.5 to 8 L, Flow: 0 to 14 L/s
Resistance
Less than 1.5 cm H2O/L/s
Cal Check
With 3-liter Syringe
Power
(2) AA alkaline batteries, capacity for 400 tests
Size and Weight
3.3 x 6.2 x 1.7 inches; 9 ounces
Optional Cradle
Interfaces spirometer to printer or PC
Features
Ultrasonic Flow Measurement
Single patient use spirette
Alpha Numeric Keypad
Graphic Curve Display
Meets ATS 1994 diagnostic standards
Meets NIOSH/OSHA and SSD guidelines
Customizable Configuration
Automated Coaching Messages and QC
Automated Interpretation
Two AA Batteries
Stores up to 700 test sessions
Optional 8.5 x 11 or A4 printout
PC Data Transfer
Toll free technical support
3 year exchange warranty
Click here for a sample report