NDD EasyOne Specifications

Test Types

Forced Vital Capacity (FVC), Flow Volume Loop (FLV), SVC, Maximum Voluntary Ventilation (MVV), OSHA Cotton Dust, Disability

Parameters

FEV1, FVC, (or FEV6), FEV1/FVC, (or FEV1/FEV6), MVV, FEF 25%-75%, PEF, FET, FIVC, PIF, variability, pre-post % change, VC, IRV, ERV, VT, IC, FEV1/VC

Predicted Normals

NHANES III, Knudson 76, Knudson 83, Morris, Crapo, Dockery, Hsu

Languages

English, Spanish and French

Accuracy

Volume: +- 3% or 0.050 L, whichever is greater
Flow: +- 5% or 0.200 L/s, whichever is greater (except PEF)
PEF: +- 10% or 0.400 L/s, whichever is greater

Range

Volume: 0.5 to 8 L, Flow: 0 to 14 L/s

Resistance

Less than 1.5 cm H2O/L/s

Cal Check

With 3-liter Syringe

Power

(2) AA alkaline batteries, capacity for 400 tests

Size and Weight

3.3 x 6.2 x 1.7 inches; 9 ounces

Optional Cradle

Interfaces spirometer to printer or PC

Features

Click here for a sample report

NDD EasyOne Spirometry